mrtodd - posted 18:24, Wednesday 09 September 2026 (12252)
Test Cavity Alignment Work

M. Richardson, M. Todd

We did a couple of tasks related to the alignment of the test cavity in efforts towards getting an error signal. We were not seeing very strong dips at REFL even with good peaks at transmission, so we tried to diagnose why that could be. Our first thought was that we needed to align the OC so that the cavity axis was set so that the promptly reflected and cavity REFL beam could overlap well at the same time as being aligned to the cavity axis. We also tried to remedy some of the clipping we were observing on the output path (looking at the TRANS camera), and were more successful with this. We still have some aperture clipping somewhere in the output path it seems but I think it is good enough for our purposes.

This did not really remedy the lack of REFL dips we were seeing, and after awhile of trying to understand what was going on, we think it could be that the expected REFL dips are just so small we weren't looking on the right scale. I think before we were tricking ourselves with larger REFL dips when we are clipping on the PD as the dips to the HG10 are similar magnitude at REFL (while much smaller at TRANS). We will try putting a little bit more power on the RFPD and zooming in to see if we get kind of dip while being well centered and also looking to see if we can get an error signal.