CoupledCavity General
mrtodd - posted 18:34, Friday 04 September 2026 (12244)
Afternoon Test Cavity Setup

M. Richardson, M. Todd

Mitchell spent some time re-doing some of the output path by replacing OM2 with a 1" optic and putting it further away and pointing to come back to the newly added OM3 which seemed to remedy the problem of clipping on the previous OM2 we were having. We then re-installed the OC and got the cavity roughly flashing, and with some alignment tweaking got it so that we were seeing clear FSR + TMS mode structure. We could really reduce the mis-alignment peaks (odds) but as expected we cannot reduce the mode-mismatch peaks (evens). Mitchell estimated the TMS be about 4MHz and the FSR to be 15MHz which seems to line up with the mode spacing seen in the picture, as we estimate the mode scan to be about 20MHz. We will try a little bit of mode matching next time and fine-tuning the alignment, and then we will try to generate a PDH error signal.