LASTI General
mrtodd - posted 18:44, Thursday 27 August 2026 (12236)
Afternoon Test Cavity Setup

M. Richardson, M. Todd

This morning we suspected that the input coupler in out setup was wedged so we rotated (rolled the mirror in the optical axis) to see if the transmitted beam deflection angle changed, and indeed it did. We reasoned that the only reason the deflection angle of the transmitted beam should change was if there was a wedge. We did not calibrate this in any way so we have no good estimate of the wedge value (but it's ballpark probably 0.5 degree), but we are convinced it is wedged in some manner.

This afternoon we tried walking the beam around with the PSL periscope motorized mounts without any success. We actually ended up losing the beam, so we decided to start over by removing the output coupler and aligning the beam to the TRANS port optics. But we could not recover the alignment to REFL by walking the beam around.

Tomorrow we will wrap a power meter in foil and see if we can measure any transmitted power through the 60ppm output coupler to hopefully get some small signal on the TRANS PD before aligning the cavity. We will also have to get the REFL beam back, so Mitchell will once again be traversing the beam tube in that effort.