LPS optical power output is getting divided unevenly by the fiber splitter. As measured at the Fiber splitter end connectors Aug 3 2026- LPS 1: 402 µW LPS 2: 278.5 µW LPS 3: 1045 µW
Thank you, Bruno. Could you please add power levels at other 5 ports of the 1x8 beam splitter?
Also, the second figure seems to show tight fibre loops (radius of a few cm) of the fibre beam splitter which is attached to the ISI via tape. The loop diameter should be, ideally, larger than 20 cm to reduce the fibre stress. Also, tapes are not ideal because they lead to crackling noise. It would be best to avoid utilising tapes with fibres.
Attached is a a diagram of the power distribution to the 3 LPS sensors. The incoming power is measured at 2.1mW which leads to a stainless steel 1 ->2 splitter (2.1mW -> 1mW+1mW.) One leg of this splitter goes to LPS #3 while the second leg goes into a second splitter ( 1mW -> 292µW + 400µW). These go into LPS #1 and#2. It is possible the tight radii of the fiber spools are causing this uneven splitting from the second splitter, but Vermeer and I unraveled some of these and saw no difference in the output power. Seems more likely to be a problem with the glass splitter. The spools are lightly held by UHV foil, not tape. Surely this fiber management can and should be done differently. We will work on this.
Very nice, thank you, Bruno. As Yulin mentioned in the email, we do not require the 1x2 splitter. Instead, all LPS should be connected to the output of the 1x8 splitter. The input of the 1x8 splitter should be connected to the feedthrough:
feedthrough -> 1x8 -> LPS 1,2,3
Regarding the fibre loops, yes, they do not signifincantly change the output power. The loops cause a separate problem with back reflections and crackling noise. We would like to avoid loops with a diameter less than 20 cm.