dorot3a - posted 09:30, Wednesday 29 July 2026 - last comment - 12:08, Friday 31 July 2026(12213)
GRAVITES 2fl setup loss tabulations

I re-measured loss on indidivual components in the 2fl interferometer on Monday. We're seeing a lot of extra loss right now within the isolation box in the red lab-- hopefully this is just due to a kink in the fiber or a loose connector and not a break inside the insulation. When I respliced this spool in May the loss was measured around 5dB (6dB including the up and down link). Losses measured in dB. We still haven't spliced the majority of components so a lot of this will be connector loss. If we can deal with the large amounts of extra loss from the spool and delay line in the upstairs loop we're on track to have around half the loss compared to the 40km setup.

  common 1 2
Spool 5 4.4 12.6
PZT -- 0.5 0.2
AOM (at 24dBm drive) 2.6 3.2 2.6
Delay line -- 1.6 6
Total from bs1 out to bs2 in 7.6 9.7 21

 

Comments related to this report
dorot3a - 12:08, Friday 31 July 2026 (12214)

We found a kink in the fiber lead from spool 2 at the feedthrough from the second isolation box which was the cause of the excess loss. The loss of spool 2 incliding the up- and downlink fibers is now down to 6.3dB, for a total loss of 14.7dB on arm 2. Also got mostly length matched across all 3 arms (to within 80cm) today-- the patch cables we added for length are currently at -1.2dB (arm 1) and -1dB (common). We should be able to get the loss lower, but the splicer here seems to really struggle with ULL fiber.