Yulin, Denis, Rich, Bruno
I visited MIT last week to test the laser position sensors (LPSs) on the ISI table inside the HAM-X chamber. The LPS is an interferometric position sensor that will later be integrated with HoQI and COBRI on the SenIT suspension table for combined readout and noise-performance testing. Since the suspension table was not ready before my final day at MIT on July 10, three LPSs were instead installed on the HAM-X ISI table with three separate fixed mirrors, as shown in Figure 1. Each sensor and mirror is mounted on a holder secured to the platform using screws or dog clamps. The distance between the front surface of the titanium sensing head and the mirror surface is set to 48 mm, which is the nominal operating distance and gives close to maximum interferometric contrast. All three sensors can reach at least 85% contrast after alignment, and some have previously reached around 90%. For future integration with the suspension table, the simplest and lowest-risk procedure is to release the clamps, move the existing LPS with its holders to the required positions, preserve the 48 mm separation, and perform a final alignment. The correct sensor height has already been included in the holder design based on the CAD model of the suspension table.
The in-vacuum system also includes the DB-25 cable, which connects the photodiodes inside the LPSs to the vacuum electrical feedthrough and then to the transimpedance amplifier outside the chamber. The in-air electronics are shown in Figure 2 and consist of the laser temperature and current driver in the upper 19-inch chassis, the transimpedance amplifier in the lower chassis, and the laser box mounted near the 1550 nm fiber feedthrough. As shown in Figures 3 and 4, the laser box is bolted to a flange adapter immediately next to the feedthrough, giving essentially zero exposed in-air fiber between the laser box and the chamber. This arrangement is beneficial for LPS displacement-noise performance and should also be compatible with a future LIGO observatory installation.
The AA and AI filters are connected to the m1sus CDS system using channels 1–4 of ADC2 and DAC3, with DAC channel 1 connected to temperature control and channel 2 connected to current control. The LPS CDS model has been built and is running on m1sus as M1LPS and the LPS_MAIN medm screen is linked to the LASTI_OVERVIEW medm screen.
Before I left MIT, the SQS 1550 nm 1×8 splitter had been shipped from Caltech but had not yet arrived. For the initial checks of the sensors, cabling, and CDS model, I therefore used the 1550 nm 1×2 splitter previously used for Arnaud’s CRS setup and connected LPS1 (H1) and LPS3 (H3). Once the 1×8 splitter arrives, it can replace the temporary splitter so that all three LPSs can be operated simultaneously. The attached PDF shows a preliminary measurement with H1 and H3 while the chamber was still at atmospheric pressure and the ISI was locked to ground. The two sensor spectra show good agreement, and the differential signal reaches below approximately 1 pm/sqrt{Hz} from a few hertz to several tens of hertz. The remaining low-frequency noise is likely influenced by incoherent air-current fluctuations, while the higher-frequency noise is likely dominated by residual motion of the ground-locked ISI. Both contributions are expected to decrease once the chamber is pumped down and the ISI is floated.
This log is related to the ongoing discussion about in-air fibres between Yulin and Rich/Alex:
In summary, I think we should avoid any sensitive component exposed to air in a precision measurement system, if possible at all. The sensors have sub-pm precision down to 20 mHz, while the optical path inside ~meter-scale in-air optical fibres fluctuates by ~1 rad on a time scale of a few sec (depending on the air currents, insulation, fibre length, etc.). This is 6-7 orders of magnitude above the level we like to measure. Even though the coupling is small and common mode, it is still visible in noise spectra as discussed below.
We spent a significant amount of time studying this problem in Birmingham. The first attached figure shows the LPS performance achieved while sensing a suspension with 3 sensors. The noise level 0.3 pm/sqHz at 0.1 Hz, a hair above the LPS sensitivity with stationary mirrors (0.2 pm/sqHz at 0.1 Hz). The laser is installed near the feedthrough in this measurement, and no fibre is exposed to air.
We then moved the laser away from the flange and installed an in-air fibre in between. As a start, we used Thorlabs SM patch cords, inspected 1 m and 5 m long options. In both cases, we see glitches straight after the installation. The glitches are caused by the stress relaxations in the fibre jacket and decay over time, but still appear from time to time, likely due to T-induced stress relaxations in the jacket and indices of expansion mismatches. The glitches and their count are shown in the second and third attached figures.
We have then tried a different set of fibres as shown in the attached figures 4-6. We got the best results with UHV SQS fibres, likely because there is no soft jacket; the fibres are insulated in a stainless steel tube. Still, none of the curves match the 0.3 pm/sqHz performance at 0.1 Hz. We tried further insulating the fibres with pipe foam. This works fine in reducing the noise of Thorlabs patch cords, but also leads to glitches because the insulation relaxes in air, similar to the jacket.
Therefore, I think we should not compromise the performance and mount the laser on the flange as Yulin has done, if possible at all.