Posting this here because I seem to forget some detail of the process every time I do this and in case it's useful to anyone in the future.
You will need: Netfinder or similar application, prologix GPIB configurator application
First: set IP address in netfinder. Power the GPIB controller and connect it to a computer with whatever method you intend to use (via router, directly to usb, etc.)
For local connection use something like 169.254.1.100 (local, ie USB connection) or 192.168.1.100 (default) for eg. a router connection (or replace 192.168.1 with your subnet if they don't match)
Subnet mask should be 255.255.0.0 for local, 255.255.255.0 for router network. Gateway blank for local, or router IP for networked. Configure the device, make sure you get an success message, and close netfinder.
Plug the controller into the instrument you're using it with. Click the device in the 'select device' box and set it to controller. For a SR785, the default address is 10. The other fields seem to auto-set after connecting to the 785 no matter how you set them on my computer, but in the case that doesn't work, the 785 uses a LF EOS. Press 'Update CONNECT.INI' and then exit.
On the 785: In system> remote: set output to GPIB, addr. 10. In output, set hard copy button > ascii dump, destination > GPIB, GPIB ctrl: host