REMINDER: When comparing to the values at the sites, recall that this OPO has the O3 mirrors
This week
532 Power at Fiber Launch: 33.5 mW
REFL 532 Power (Unlocked): 21.0mW
REFL 532 Power (Locked): 7.17mW
Trans 532 Signal (Locked): 8.13 (i think this is in uW but seems low, need to check calibration)
532 Mode matching = 0.78
Seed Incident: 32 uW
Seed Trans (no gain): 0.150 V
Seed Trans (max parametric gain): 2.25V
NLG: 15.0
Optimal OPO Crystal Temp: 34.967 C
Threshold power (in TEM00 mode) ~29.7 mW incident. Roughly 29% higher than at the start of our test
Fiber efficiency was about 65% when i measured with power meter. Best I've seen is 67% with good alignment. I think no clear degradation
I'm seeing the opo demod phase drift around on me over time which i suspect is causing the slow anti-correlated fringes in the incident and transmitted power of the OPO im seeing when locked. I may have a bad cable somewhere in the OPO RF chain. Will track down next week if I have time. Not a major impact on the test.
Comparing this to the behavior of the crystals at the sites, they typically see a similar level of degradation after only a couple of days. https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=82196
This initial rapid degradation is followed by a slower decline in performance over the next year: https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=82080
Our initial degradation seems to be occurring less rapidly. Hopefully this remains true for the long-term degradation.