These measurements of the servo loops were taken in mid-June immediately before I left Vienna. Since then the performance has degraded a lot, it's unclear why-- maybe something to do with a power outage that happened a few weeks ago. Posting these still for reference.
Servo 1 (blue/downstairs ifo): predictably, this loop is much cleaner as it's contained entirely in the downstairs lab isolation room. UGF a little below 20kHz, feature from pzt resonance around 5kHz, decent amount of phase margin. Crossover just below 300Hz, but quite a lot of excess noise above 200Hz, and more phase noise than I'd like to see below 1kHz.
Servo 2 (yellow/2fl ifo): Also predictably, much noisier. UGF around 10kHz and also decent phase margin but with additional noise below 5kHz. Crossover around 300kHz, generally noisy at low frequencies (also predictably) with a lot of noise below 100 Hz, probably coming from the fiber link-- this was also visible in a power spectrum of the error signal taken around the same time, especially between 30-50Hz.
Note that on servo 2 I changed R7 3.3k -> 887 Ohm in order to get the crossover below 300Hz when initially installed in May.
These were good enough to lock at the time at least long enough for Tom to perform checks with the photons while I was away so they were left as is.