SQZ General
eoelker - posted 17:33, Thursday 14 May 2026 - last comment - 18:55, Saturday 16 May 2026(12183)
NLG Log: Week 3

REMINDER:  When comparing to the values at the sites, recall that this OPO has the O3 mirrors

 

This week

532 Power at Fiber Launch: 29.85 mW

REFL 532 Power (Unlocked): 18.7 mW

REFL 532 Power (Locked): 6.43mW

Trans 532 Signal (Locked): 8.17 (i think this is in uW but seems low, need to check calibration)

532 Mode matching = 0.78

 

Seed Incident:  35 uW

Seed Trans (no gain):  0.163 V

Seed Trans (max parametric gain): 2.57V

NLG:  15.8

Optimal OPO Crystal Temp:  35.029 C

Threshold power (in TEM00 mode) ~26 mW incident
 
Notes:
We're seeing our first evidence that the crystal is slowly degrading.  Higher pump power now required.
Fiber efficiency is a bit lower.  Haven't tried optimising alignment yet.
Comments related to this report
eoelker - 18:55, Saturday 16 May 2026 (12184)

I was looking through some old aLog posts trying to compare the threshold level we've seen so far to the old PPKTP crystals.  Looks like its reasonably comparable to what Maggie had shortly after a crystal move during O3 comissioning:

https://alog.ligo-la.caltech.edu/aLOG/index.php?callRep=41060

23 mW threshold.  Which is about where we started before the currently observed aging

 

I'm starting to think that the aging in this test might be a bit hard to compare to what we've seen at the sites.  I'm using the O3 mirrors (which require higher circulating 532 nm power for a given NLG), but I'm using the O4 nonlinear gain level.  Given that I don't plan to upgrade the mirrors on this OPO in the near future, perhaps I should reduce the NLG to the level used for O3 so its a bit easier to compare to our old data.  I'm going to chat with folks at the SQZ meeting next week.