Bruno - posted 11:30, Wednesday 13 May 2026 (12180)
Vacuum Bake Oven 05_06_26 RGA scans
Took RGA on new chamber after 150 C bake for 16 hours. RGA+Oven (no calibrated Argon) UHV angle valve must be closed. Green Neupro valve is insufficient. 

Strong smell during bakeout - Foil tape adhesive on the heater trace? 
Will have to redo the insulation / take off yellow Neoprene Coolskin for high temp bakeouts. 

Prior to bakeout Helium leak checked all cf flanges using the RGA in leak test mode. Absolutely no leaks detected though there was a small upward drift when I bagged up the second flange on the RGA tree. This might be from the electrical feedthrough. 