[Eric, Daniel]
Today, we tuned up the co-resonance on the LASTI VOPO. For our measurements of rough NLG, we used the following seed power and pump power:
Baseline Seed Input Power: 60 uW
Pump Power (measured from OPO REFL): 9mW
The crystal stage was re-referenced and we found all the spot locations roughly as follows:
Crystal Bottom Edge: -1.9mm on pico stage readout after negative reference (DO NOT POSITIVE REFERENCE!)
Crystal Top Edge: 2.86 mm
| Spot | Position (mm) | Temp (C) | Nonlinear Gain |
| 1 | -1.215 | 33.75 | 2.6 |
| 2 | -0.773 | 33.75 | 2.4 |
| 3 | -0.331 | 33.75 | 2.5 |
| 4 | 0.115 | 33.75 | 2.0 |
| 5 | 0.555 | 33.75 | 2.9 |
| 6 | 0.997 | 33.75 | 2.9 |
| 7 | 1.435 | 33.75 | 3.1 |
| 8 | 1.874 | 33.75 | 3.0 |
| 9 | 2.319 | 33.75 | 3.1 |
| 10 | 2.744 | 33.75 | 3 |
After optimizing around spot #9, the target co-resonance parameters while locked were:
Temp = 35.372 C
Stage Position = 2.488 mm
Pump Power = 9 mW
Nonlinear Gain = ~5
After this, we disabled the stage and unplugged the stage controller from the chamber to prevent additional heating of the oven parts.
Eric also added an RF preamplifier to the CLF-2F beat signal from the CLF PD before the demodulator. This improved the beat-tone amplitude enough to resolve it on the logarithmic power detector built into the demodulator (needed for the guardian to observe co-resonance)