Rich, Bruno, Arnaud [remote] CRS1 Rich added .98g of washers (as per https://dcc.ligo.org/LIGO-T2600096) to the bottom 8-32 screw of CRS1 to get the COM below the flexure point (to address the issue described here: https://alog.ligo-la.caltech.edu/SEI/index.php?callRep=2564). We measured a new resonant frequency of 22.8mHz. CRS2 The adjustment mass was stuck on one side of the mount. It was recentered with the proof mass on its lock. Rich mentioned the it was slipping otherwise. Bruno checked the TEC controller, and the 'open' red led was on, we will need to investigate. For both CRS, the light levels on the pds went down on March 1, and had not recovered since then (-400cts instead of a few thousands). Since the preamp is shared between the two, Bruno power cycled it (D2300154-v4), and the signals came back. Balancing will continue tomorrow.