Frosti-Lite General
Evan Hall, Bruno Becher, Nelushi Vithanachchi - posted 16:16, Sunday 29 March 2026 (12167)
FROSTI-lite scan results at different distances
We did a series of scans at different distances of the thermophile from frosti.

Test 1- March 12th 
With the  ZnSe lens 
Power supply 
Hylec (Drever and weiss)
20V ,1.448 A and 28.9W
Agilent (Barish and Thorne)
19.98V, 1.463 A
Drever and Weiss get warmer than Barish and Thorne. Though there is no apparent asymmetry in the thermophile scans.
Total Ohmic power= 20*1.448+19.98*1.463= 58.19W
The total power captured by the thermophile is calculated as follows.
P(mW)= The power recorded by the thermophile
A=10*10 mm^2 (Area of thermophile)
The scan step size= 6mm
The total power =sum( P*10^(-3)/(A*d*d))

Test 2- March 13th
Without the ZnSe lens
The scans look asymmetric (The thermophile was likely angled- Will retake the data)
In order to reduce the disparity among top and bottom half, the power supplies were set to different values
Power supply 
Hylec (Drever and weiss)
17.00 V ,1.232 A
Agilent (Barish and Thorne)
19.98V, 1.465A
Total Ohmic power= 50.2147 W

The results are attached. 