[Eric, Daniel]
Last Friday, we attempted to fire up the squeezer in LASTI for the first time since we pumped down at the end of last year. Initial efforts to do this at the beginning of the new year were somewhat stymied by the difficulty in controlling the OPO oven in the chamber. After getting the old LIGO-style TEC driver working (along with the MEDM screens for it), this was resolved last week in this alog. The TLDR is that we were able to get all the subsystems of the squeezer back on line with the OPO in the chamber pumped down to ~1e-6 torr. We had to quickly re-align the homodyne detector to ~95% visibility (not much time was spent optimizing this for a quick check) and we were able to measure squeezing levels just over 6dB.
As we continue to get the slow controls automation back up, we also documented the state of all the CMBs as an initial baseline (with rough transfer functions). No transfer functions for the OPO or SHG PDH loops were taken (these seemed nominally okay... OPO CMB could possibly use a tune). These are tabulated here:
| Setting | Nominal Value |
| Input Gain | 14 dB |
| Common Path Boosts | Boost 1, Boost 2 |
| Slow Path? | No |
| Setting | Nominal Value |
| Input Gain | 9 dB |
| Common Path Boosts | Boost 1 |
| Slow Path? | No |
| Setting | Nominal Value |
| Loop Gain | 16 dB (on demod monitor channel) |
| Boost Filters | Boost 1, Boost 2 |
The LO PMC OLTF was taken while the turbo pump was running on the chamber. From this, a noticable 1 kHz spur ~10x larger in ASD than the background noise (measured by Anders here) was noted in the error signal voltage (from the demodulator)
| Setting | Nominal Value |
| Input Gain | 31 dB |
| Common Path Boosts | No |
| Slow Path? | No |
| Setting | Nominal Value |
| Input Gain | 6 dB |
| Common Path Boosts | Boost 1, Boost 2 |
| Slow Path? | Yes, All Filters Disabled |
It became very apparent that the CLF and LO squeezing angle stabilization loops both need some retuning to be more stable. As of now, both loops have lousy phase margins (the LO loop likely has an unecessary integrator from when this was controlled with a PZT instead of the VCO method used now) and don't reliably lock with the added phase noise from the turbo pump. We could just turn the pump off for measurements, but with some loop retuning we may be able to avoid this.