Bruno set a temperature datalogger for the VEA next to the HAMX chamber to help us understand drifts in CRS1. It is plugged to the laptop between the chamber and the rack (see picture below). The software is Elitech, and is set to record datapoints every 15minutes (to see those points on the graph, you must click on the 'Reload Data' button).
Currently reading 24.3C, 15.9RH.
Attaching here data collected so far. The left Y-axis on the picture is temperature.
Plotted against the CRS drift (note CRS was railed on the first few days). We can see the correlation between the drift and the 2 degC temperature swing around Feb 25.