[Bruno, Arnaud]
On Friday, Bruno helped me recover the NKT laser for CRS1 (see log 2560).
[Rich, Michael, Arnaud]
*CRS1 damping, medm and iso script
We commissioned the CRS1 damping using ISI actuators (in lieu of the ESD damping) to reduce the ring down wait time. The damping can be ran via the CRSRX_BLND filter path (FM2) and either injected at the error point of the closed loop (with a gain of 1), or in the open loop state with a gain of -0.01. The filters can be accessed via the 6D blend filter that were updated to include the CRS RX filterbank (see the new rx path on the bottom left in the medm screen attached).
Also, a python script was written to turn on the iso loops (faster than the matlab does). It assumes the ISI damping is on. The script can be found under ~/Desktop/CRS/script/turn_on_iso.py
*CRS calibration cross check
Today we took some data comparing the CRS1 and the other ISI channels in the damped and iso state to verify calibration above 0.1Hz. The attached pdf shows the rx cps, gs13 and crs1 calibrated channels in the off state (page 1), damped state (page 2) and iso state (page 3). The dtt template is under ~/Desktop/CRS/dtt_templates/dtt_hamx.xml
OFF and DAMPED state :
ISO STATE (all loops on, 250mHz CPS-GS13 blend, no sensor correction)
*CRS1 drift
The CRS has been drifting over long periods. We observe a 12h cycle which is likely temperature induced, and a longer trend which might be vacuum induced. Between Jan 31 and Feb 1 it drifted by roughly 7.5mrad (see attached), and has been turning around since then.
*CRS1 TFs
Low resolution TFs will be running overnight to extract the CRS response.
The CRS1 response was measured, with results and fits posted here.