SQZ General
alossius - posted 21:43, Friday 30 January 2026 (12133)
LASTI SQZ reboot since pump-down, OPO mode-matching and PMC PDH transfer function and noise spectrum

[Daniel, Eric, Anders]

Last week (week of Jan 19, 2026), we resumed work on the LASTI SQZR.

OPO Vacuum Chamber

The VOPO chamber has been under vacuum since Eric and Daniel pumped it down on Nov 20th (see alog). On Thursday (the 22nd), the pressure in the chamber was aroud 5e-7 mbar when we got into the lab. Eric disconnected the pump from the chamber, and we observed the rate at which the pressure increased. By the end of thursday, the chamber pressure had increased to something on the order of 10^-5 mbar. By the end of the day the next day (Friday), the chamber pressure was around 2.5e-4 mbar. This pressure rise indicates the presence of imperfections (somewhere) in the seal, on top of the out-gassing effects. However, with the pressure staying above 10^-3 mbar over the course of 48+ hours, we know now that we can re-pump down the chamber every few days or so to maintain an ideal operating pressure. 

SQZR

It had been a while since the LASTI squeezer has been operational - not since the VOPO chamber was pumped down. We went through the standard procedure of turning everything on (locking SHG, OPO, PMC loops) and were pleased to see that there had been no significant misalignement. We did however, notice that the seed power present on sqz+readout table(s) was very low, which we traced back to a misalignment issue with the seed fiber coupler on the main laser (1064nm) table. This was quicly fixed and everything was back to operation. 

Daniel and I did a quick check of the seed and pump mode matching into the OPO cavity. The results are tabulated below, and the cavity scans can be seen here.

  Power in Fundamental (V) Total Power (V) Mode Alignment (P_Fund/P_Tot)*100
Seed (1064nm) 1.66 2.135 77.7%
Pump (532nm) 1.60 2.026 79%

*** These mode matching values are on-par with what has been measured previously (75.6% for seed/clf (1064nm), 77.4% for pump (532nm)). 

OPO Crystal Temp Control

On a seperate note, we were unable to get the Newport temperature servo for the OPO to lock without oscillating. We will need to set up the LIGO temperature controller again now that we're operating under vacuum.

PMC Loop & Controls

On Friday (23rd), we took spectra of the PMC loop's error signal and an open-loop transfer function (OLTF) using the SR785. The objective was to get a plot of the calibrated, free-running and rms error in the PMC loop. To do this, we needed to convert the y-axis of the noise spectrum to units of meters/sqrtHz and use the OLTF to divide out the loop suppression and get the free noise. To get the proper units for the noise spectra, we measured the PMC's FSR and FWHM using the DC transmission PD to get a conversion from time to meter, and measured the slope of the PMC error signal (in the linear regime near resonance) to get the conversion between volts and time. We fit the OLTF to a model of the TF that Eric had. The summary of this process is outlined below:

* The FSR was measured to be ~449ms, however, scanning the cavity with the PZT does not give us a clean (full) FSR, so this value is most likely imprecise. To get a better reading, we should scan the cavity using the laser's frequency. 

* The linewidth (FWHM) was measured 4 seperate times - each measurement was pretty different: 200, 542, 310 and 788 us - unclear to me at this point why these measuremetns differ so much. 

* Using these two measurements, the PMC cavity (p-pol) finesse can be inferred, however, the significant deviations in linewidths results in equally different values for the finesse: 2245, 828, 1448, 570. Furthermore, none of these finesse values land within 50% of the documented cavity finesse in the PMC DCC document from [year] (see LIGO-T2300300-v5 - link unavailible at this time as I don't have DCC access yet), which is 220. 

* Due to this inconsistency, the cavity finesse was taken to be the 220 value from the DCC document for the remainder of the analysis. 

* To get a conversion between time and meters, we use the cavity finesse and laser wavelength to get the value of the linewidth in meters, and compare this value to the measured linewidths in time (*which I'd like to remeasure soon*). 

* To get a conversion between meters and volts, we use the slope of the error signal with the x-axis converted to units of length using the time-to-meter conversion just mentioned. This slope then has units of volts/meter and gives us this conversion. The data collected and used for this conversion can be seen in the table below:

dY [mV] dX [us] dX [nm] (multiply by time-meter conversion)
768 42 44.1
578.25 98 1.03          *clearly an outlier*
486 48 50.4

* The uncalibrated noise spectrum of the error signal was recorded using the SR785 FFT function. By dividing the y-axis (nominally in units of Vrms/sqrtHz) by this slope (volts/meter), we obtain the calibrated error spectrum in units of meters/sqrtHz. A plot of this calibrated noise spectrum can be seen here.

* The open loop transfer function (measured using the SR785 via the excitation point on Eric's servo) was taken and plotted on top of the transfer function model. In order for the measured TF to match the model, the model's overall gain parameter was bumped up from 25 to 42 dB. A plot of the measured and model TFs can be see here. At the point of this measurement, the UGF was around 1.4kHz, however a few days earlier the loop was stable at a higher gain (~16dB vs this at 4dB) which corresponded to a higher UGF of ~4kHz - we do not fully understand why the loop's lockpoint changed over the few days between locks. 

* Using the model for the TF, we computed the free-running error in the system, by multiplying the calibrated error with |1-OLTF|. We then computed the running rms error from high to low frequencies, and plotted the three curves here.

This data will serve useful for further loop tuning. 


 