Displaying report 1-1 of 1.
Reports until 11:31, Friday 16 January 2026
General
richard - posted 11:31, Friday 16 January 2026 (12132)
CRS Noise Spectra

We took some noise spectra of the CRS with and without sensor correction on, see attached. The CRS is exceeding the ~prad/sqrt(Hz) goal above 20 Hz.

With sensor correction on, the noise roughly matches the T240 noise from 0.1 - 20 Hz and is dominated by residual motion coupling through nonlinearities of the readout. The common signal (i.e. rotation) supression in the noise estimates is typically limits to roughly two orders of magntitude. The signal leaks into the noise due to nonlinearities in the readouts. This is why sensor correction decreases the noise in the CRS. We expect this to decrease even further when we put the sensor in loop. Below 0.1 Hz, the CRS is well below the seismometer noises. The CRS noise in these plots is just the readout noise so it's expected to be less than the goal noise below 0.5 Hz. Below 0.5 Hz, the goal is dominated by the thermal noise of the CRS which would only show up in the angular readout.

Non-image files attached to this report
Displaying report 1-1 of 1.