As mentioned in log 12121, Amit and George have rotated the C-6D top plate to compensate for the RZ offset (discussed in 11846) that we have introduced back in Dec 2024 likely because we overcompensated for the air currents. This time we noticed only a small change in the RZ control signal in air (when the chamber curtains were closed) and after the pump down (150 cnts with 0 coil driver impedance = 0.28 uN m of torque that is likely caused by T change or buoyancy coupling via other degrees of freedom). We still have two items to fix / improve: * install thermal shields * replace the fibre to remove the RX/RY assymmetry Still, I decided to look into the C-6D noise budget more carefully and compare its noise with BIS' (log 12123) and T240's sensitivity. The noise budget is shown in the first attached figure: the red trace shows the raw RY signal and the green trace shows the calibrated spectrum. At low frequencies, the measured motion is limited by the classical temperature noise. The actuator (DAC) noise and thermal noise are significantly lower than the classical temperature noise in this state (without the T shield). C-6D measures the platform tilt from 7 mHz. I tried to assess the C-6D own noise by subtracting the tilt spectrum measured by the X degree of freedom from RY. The result is shown by the magenta trace. Above 30 mHz, we cannot assess the C-6D motion because X is limited by the translational motion of the platform. There is likely a bit of miscalibration at 17 mHz (the RY eigen mode) and the C-6D resonance is still seen. I may try to correct this. The second attached figure shows the C-6D RX, RY estimated noise spectra (RX - Y/g and RY - X/g) and their comparison with BIS and T240. Similar to RY, RX is likely a bit miscalibrated around its eigen mode of 9 mHz. In the 7-40 mHz band (where I think the tilt coupling is most critical), C-6D has a factor of 4-5 lower noise compared to BIS and a factor of 15-20 lower noise compared to T240's noise. The next hardware update should likely be the installation of the thermal shield.