The platform is currently stabilised with BIS in loop. The blending frequencies are X: 40 mHz, Y: 80 mHz, Z: 20 mHz, RX, RZ: 67 mHz, RY: 32 mHz. I think the X/RY blends are a bit conservative with some room for improvement but I tried to avoid ringing the blends during the first iteration. The Y/RX blend is a factor of 2 higher than Y/RX because we have one of the station 2 coils shorted relative to the ISI that likely couples extra noise (as already mentioned in 12122). The RZ blend is also a bit higher than RY because the seismometers are tuned to couple 10% of the horizontal motion. The Z blend can be as low as 5-10 mHz from the noise figures but I decided to start with a more conservative value. I also blend BIS with GS13 at 0.3 Hz. By design and from the noise figures, we should blend at 0.7 Hz. However, we had to set the seismometer force feedback UGF at 1.5 Hz vs the design value of 4 Hz because we excited the first parasitic mechanical mode of the proof mass at 35-40 Hz during the pump down. We will need to retune the low-pass of the analog servo and then shift the force feedback to 4 Hz in the future. Further analysis is needed but at a first glance it seems that BIS / GS13 blending at 0.3 Hz does not couple much GS13 noise at low-frequencies. The sensor correction is currently off because the ground STS-2 seismometer seems to be turned off. I have retuned the notches in the ISI isolation filters because we run up new mechanical modes around 84 Hz and around 125 Hz. It seems that the new sensors (BIS, CRS) have not added any low-frequency parasitic mechanical modes. The attached figures show the spectra of the out-of-loop T240 sensor. Since STS-2 seems to be off, I had to take the reference trace a couple of hours in advance and the seismic spectrum might have changed a bit. However, overall, I think the spectra make sense: we should achieve a suppression of a factor of 30 at microseism for Z, a factor of 10 for X, and a factor of 2-3 for Y. There is also an enhancement in noise below 100 mHz that may or may not be related to the elevated tilt during the platform stabilisation. Will need to assess this in the future. I also think that T240 is rotated by 0.2 rad relative to the ISI X/Y axes and mixes our basis a bit. Not a big deal for now but might be good to rotate T240 in the future. I think these slower blends compared to the GS13 ones (100 - 250 mHz) look promising and should be able to improve the suppression of the vertex motion and reduce MICH / SRCL / vertex WFS noise in DARM below 25 Hz. Similar to Hang Yu's 2017 5-Hz paper and similar to the C-6D's paper, the propagation of the platform motion via the suspension and the interferometer's global control system is required (and ongoing) to qualitatively assess the improvement.