As noted in log 12121, the Birmingham / Tsinghua team has installed our 1D sensometers inside the HAM spring hutches. We call the seismometers BIS as a short of bison because the seismometers look like these horned fellows (at least we tried to design them this way). This abbreviation also stands for the "Birmingham Inertial Sensor".
Our goals were the following:
* build seismometers with a 5-10 improved noise compared to T240 in the 10 - 100 mHz band (required for optimal ISI stabilisation). It seems that all modern Trilliums, Guralps, STS-2 follow the Wielandt's approach and rely on leaf springs where antispring is created by the spring's shear force. We followed a similar approach with lots of smaller and one key difference:
* make them UHV and fit inside the spring hutches. As far as I know, T240 cannot fit in these places. UHV environment provides both opportunities and challenges. The key opportunity is that air damping in the small gap between the capacitive plates limits the thermal noise of the state-of-the-art commercial sensors. The key challenge is that it is difficult to utilise capacitive sensing in vacuum with picometer resolution because the amplifier must be near the seismometers in vacuum.
In summary and so far, the seismometers managed to achieve a factor of 3-4 improved sensitivity compared to the T240 seismometers and a factor of 2 worse than their design sensitivity. This will be investigated during the commissioning period. The first attached figure shows the noise level from Dec 24 in RY degree of freedom. The ISI tilt has been selected because this is the quitest angular degree of freedom and may reach down to the seismometer own noise:
* The red and gray trace compare BIS and T240 Z acceleration. There is some extra noise in T240 above its self-noise and above BIS. This might be due to magnetic coupling or due to imperfect diagonalisation of the seismometer.
* The green and black traces show the inferred ISI RY tilt acceleration from the two vertical BIS signals and from T240 X signal (that couples tilt with a usual g/omega^2 coefficient). The figure shows that we measure the ISI tilt above 25 mHz and sit on some other noise at lower frequency. The noise is above the design and will be further investigated.
* The cyan trace attempts to reach the BIS's self noise by subtracting the ISI tilt measured by T240 X. We reach acceleration of 0.2 nrad /s^2 / sqHz at 25-50 mHz. This is a factor of approximately 2 above the design shown by the orange trace.
* The blue and magenta curves show the T240 self noise and GS13 self noise for comparison.
The second attached figure shows the same as above but in units of rad/sqHz.
As mentioned in log 12121, we have a cabling problem with station 2: one of the coil wires is shorted relative to the ISI. This creates a ground loop in the system and Lari estimated that 1/4 of the coil current returns via the ISI ground. This likely couples noise from the ISI ground to the seismometer. Indeed, RX seems noiser than RY because RX forms by all three vertical signals.
This noise is shown in the third attached figure: RX vs RY. The figure also shows raw seismometer signals that are tuned to S_1/2 = Z +/- 0.1 H. The H-coupling coefficient is tuned as a compromise between the horizontal sensitivity and seismometer diagonalisation (we require more than 60 dB subtraction of horizontal signals from the vertical ones to reach the design sensitivity).