GRAVITES General
begum - posted 00:33, Monday 01 December 2025 (12114)
GRAVITES data acqusition set up

I set up the Keithley DAQ6510 for long term data acquisition for the GRAVITES setup. 
I attempted taking low frequency data with SR785 last week, setting the bandwidth to 1 Hz, with 100 points, 10 averages. It was not able to complete this set even once (the device channels occasionally saturates and this returns the average number to 0). Of course for long term studies on the phase offset between the control field and the signal field, we need to be able to record long data segments and post-process ourselves for removing glitches and proper averaging. With the pzt drfit issue solved, the only thing left to do for long term monitoring is setting up the data acqusition. 

DAQ6510 takes data from the front or the back terminal. The back terminal channels 101 and 103 are set-up for monitoring the temperature of the laser prestabilization setup. Channel 110 is for PZT mon, 111 is for the 1545nm HD and 115 is for the 1550 nm HD tap-off we usually use for monitoring on the oscilloscope. 110, 111 and 115 are set to acquire DC coupled voltage data. In the current set-up, I aimed to still monitor PZT mon and 1545 nm error signal on the oscilloscope at the same time as the data acqusition, so there are BNC T-splitters on the path. PZT mon is split once, 1550 nm HD split once, 1545 nm HD is split twice (the first split is between the error signal to the servo and the diagnostic arm, then the diagnostic is also split into two: oscilloscope & DAQ6510). The BNC cables going to the back terminal of the DAQ6510 are labeled clearly, so the oscilloscope can be disconnected while acquiring data. For reference, input impedance of the DAQ6510 is >1 MΩ.

The device is connected to the local network with an ethernet cable, Menu --> Connection --> LAN menu shows the IP address.  On a computer on the same network, one can access the machine remotely by typing in the IP address on the browser bar. Attached is a screenshot of such a connection on the GreenLab desktop, showing the Home screen, Channel 110. The "Extract Data" layer shown on the lefthand menu allows us to download data to the computer directly from the designated buffers. Alternatively, one can click on "Save to USB" on the Home screen for immediately saving the buffer data to the USB (plugged in physically to the DAQ6510 front left USB port), or automatic saving to the USB can be arranged while setting up the measurement itself. Buffers don't auto-clean themselves, so the data from the previous scan is intact (and download-able) until a new scan is started.

Menu --> Scan is the main menu that lets us select the channel numbers, the channel type, where and how often to save the data. The device first saves the data to a buffer, then to the usb. Both Buffer 1 & 2 have a capacity of 100 000 data points, either one can be selected from the Scan settings. The minimum Scan to Scan interval (under Scan settings) is 8 us. For 10 us interval, 2000 scans with 5 channels take 4 minutes 45 seconds, so data acqusition takes 28.5 ms, nearly all of the time spent in saving the data. This is a bit shorter if only one channel is being saved, 17.2 ms per step. Saving to usb every scan vs at the end of all scans doesn't seem to make any difference. I defined a new "buffer4" with 6 000 000 data points, with 10 us interval, this comes up to 1 day 4 hour 43 minutes of data (with a single channel). I am leaving a scan with these setting over night, for a dummy data acquisition campaign with the PZT mon signal. 