11/11/25 - Picture of damaged knife edge on rear flange with a few 47 Micron diameter wires for scale. The indent is maybe 20-30 Micron? Polished the scratch with circumferential motions of a Novaculite Arkansas stone followed by fine emory cloth with isopropanol for "wet sanding". After this I cleaned the knife edge with isopropanol and placed the flange on the chamber with a silver plated 13.25" gasket. Cleaned all open flanges with isopropanol. Torqued the bolts starting at 8 ft*lbs up to 32 ft*lbs in a star pattern. 11/12/25 - Placed a blank on the 4.5" CF flange and connected the Varian helium leak detector to the 2.75" CF port. Unfortunately still detected a leak coming from the same rear flange despite new gasket and polishing. Gerardo Moreno and Jordan Vanosky at Hanford miraculously found a 13.25" flange with a 2.75" port and 4.5" that should arrive here next week.