Reports until 18:57, Wednesday 12 November 2025
Frosti-Lite General
Evan Hall, Bruno Brecher - posted 18:57, Wednesday 12 November 2025 - last comment - 09:43, Thursday 04 December 2025(12104)
FROSTI-LITE Assemby and Initial test scans
FROSTI-LITE ASSEMBLY
Parts
4*Heating segments (Al)
4* upper segments (Al)
4* Thermistor (4022-ATH100K1R25TB-ND)
Nichrome wire (26 AWG )
-We wired the heating segments with Nichrome wire, with about 6 inches extruding out on either side
 -Tweezers can be used to guide the nichrome wire while winding however we found that using fingers to hold the wire against the heating segment was much easier
 -While wiring it is important to make sure that the Nichrome wire is held in tension against the walls of the heating segment
 -Future recommendations: For even resistance across all segments, try to make the length of Nichrome wire used in all segments roughly the same (And ofc, use the same gauge wiring)
-We then assembled the 4 heating segments along with a thermistor for each
 -We used insulation to cover the Cu leads of the thermistor. Due to space constraints in the hole meant for the thermistor we flattened the insulation beforehand. Once we were able to see the thermistor through the screw holes we screwed it in.
 -We soldered additional wire to the Cu leads of the thermistors
-Before assembling the heating segments onto the reflector, we covered it with styrofoam to prevent scratches on the polished Al surface.
 - Important: There are different sizes of screw involved: So a good practice would be to label the screw drivers and keep track of which size they are
 -Assembly of the 4 segments onto the reflector was challenging, due to the space constraints in inserting a screw driver/keys
 -New standoffs that will make this process much easier were designed by Bruno and is on the way
 -We then the 4 segments on to the back of the reflector
-Then we moved this to the orange lab , put the top cover of the reflector on and assembled it on to the stand . (Again, keeping track of screw drivers, will make your life easier)
-We labelled the 4 quadrants of FROSTI-LITE : Weiss,Thorne,Barish and Drever

FROSTI-LITE Power supply
-Each segment has its own  PCB power supply, but they’re all connected to the same main power source . All 4 PCB’s are at constant voltage
 -Bruno made a 3D printed box to house these PCBs 
-We also have a voltage divider circuit to measure the temperature of each thermistor
 -We give a known voltage across both resistors, where one resistor is known and other is the thermistor, from the voltage reading of the middle of the two resistors you can find the resistance of the thermistor which can be translated to a temperature using the data sheet 
 -Bruno also made 3D printed host boxes for these circuits as well
-In order to extend the Nichrome wire from the heating segments to the power supply we attached more Nichrome wire via crimps (Because soldering Nichrome wire is hard). Bruno made a 3D printed box to host these crimps (we used peak screw terminals) 
-Finally we also used a teflon tube to make all the leads go through (this was to make it less messy)

Translation stage Scanner code
-The gui required to operate the thorlabs translation stage is :https://git.ligo.org/tcs/frosti-scanner
  -We used the gui in: lts300_apt -> scanner_gui.py
-Follow the steps in the git repo and install the thorlabs kinesis software, NI-VISA  along with a software for the PM100 power meter
-The scanner_gui code has pre written COM ports for the LTS stages: Change that to the COM ports(of the LTS stages ) identified by your computer
 -If your on a windows and your computer does not give a COM port for the LTS stage, you can enable a virtual com port as follows
Find the controller device under “USB controllers”, it might be called “Thorlabs APT controller” or something similar
Right click-> Properties-> Advanced tab and check the “Load VCP” box
“OK” out of the dialog box
Unplug and replug the USB connection and a new COM device should appear
-In addition, if the scanner_gui does not recognize the power meter :switch Thorlabs PM100D mode from TLPM to PM100D NI-VISA with Thorlabs Driver Switcher (This can be done using the power meter software from thorlabs)
-We also adjusted the positions of some buttons on the gui, by manually editing the code 

Thermopile (S401C - Thermal Power Sensor Head, Surface Absorber, 0.19 - 20 µm, 10 µW - 1 W, Ř10 mm )
-We removed the housing of the thermopile to prevent clipping from FROSTI-LITE signal
-A IR filter (T-WG71050-E4 - Ř1" ZnSe Broadband Precision Window, AR Coated: 2 - 13 Μm ) was added on top of the thermopile sensor array

FROSTI-LITE Initial tests
Initial test scans from FROSTI-LITE, show that the radius of the annulus where intensity is maximum is around ~92 mm, which is close to numbers predicted by cao
The total radiant power is ~14  W, for around 50W of input power (all segments combined)



Images attached to this report
Comments related to this report
Bruno - 09:43, Thursday 04 December 2025 (12115)
The Thermopile detector plane was 58mm from the curved edge of the reflector housing.