Bruno Becher - posted 12:57, Wednesday 05 November 2025 (12098)
Vacuum Bake Oven report
The Vacuum Bake Oven in the orange lab is an Ultra High Vacuum system to clean and bake components to be used in Lasti and potentially sent to the sites.

9/3/25 - Rich, Slawec, and I disassembled the vacuum system and replaced the 13.25” copper gasket on the rear ConFlat flange of the main chamber where the primary leak was detected. The knife edge on both faces of cf flange was inspected. Tightened flange back on with a torque wrench set at 18ft*lbs tightening every 13th bolt, going up 4ft*lbs each time the bolts refused to twist. Ending at a torque of 34 ft*lbs. We also replaced the copper gaskets on Valve 4 to the main turbo, and the RGA flange where leaks were also detected. All internal surfaces were wiped with isopropanol. We also took apart the Forelines and cleaned the Viton seals in the hand crank valves. 

9/7/25 - Air leak in system at back flange still detected with helium though less intense than before gasket replacements.  

10/19/25 - RGA scans showed a usable Hydrocarbon signature at 1.5e-11 Torr*L/s despite the leak. 

10/29/25 - After a 24 hour 150°C empty bake the leak "opened up" as shown in the attached RGA scans. Helium detector showed leak is coming from same rear chamber ConFlat flange.

11/4/25 - Disassembled the Vacuum system to thoroughly inspect knife edges on rear flange and perhaps replace 13.25" 275CF CU gasket.