Today we were able to recover a decent escape efficiency on the LASTI VOPO with the SPPKTP crystal #3. We did a quick measurement to confirm Daniels numbers from friday and were still seeing a low value. After scanning the crystal a bit and finding that the escape efficiency was fairly uniform across the crystal (except for a few spots where we believe that the crystal is chipped), we suspected that we were hitting the bad spot on M3 and decided to realign the cavity.
I was able to recover an alignment that looked comparable to what I had when I first aligned the cavity a month ago by repeating my standard alignment procedure. The cavity beam is still tilted because, as I noted then, M3 and M4 do not have the vertical adjustability found in the site version of the VOPO tombstones, so the crystal is tilted a bit to compensate. Adjusting the vertical position of M4 is usually the final step of my alignment procedure.
We measured (1064 nm):
6.4 V on resonance
5.8 V off resonance
97% mode matching
This corresponds to a 90.6% reflection on resonance, or about 1800 ppm of excess loss, or an escape efficiency of around 97.7%. This roughly agrees with the loss we expect based on Daniel's measurement of the AR coatings for Crystal #3 (~1800 ppm according to this alog)
Since we cant do any better than this loss-wise, we're going to go ahead and reinstall the cavity in the chamber.