This week we continued reinstalling the "better" SPPKTP crystal (#3) into our cavity in LASTI. We hit a snag earlier in the week with the keepers for the crystal, but that has been resolved and we are seeing flashing in the cavity once more. After realignment of the red/green input beams, we achieve the following mode matching as witnessed by a cavity scan.
| Mode | Green Trans Voltage | Red Trans Voltage |
| TEM00 | 278 mV | 4.64 V |
| TEM01 | 72 mV | 92 mV |
| TEM10 | 72 mV | 84 mV |
| TEM02/20 | 110 mV | 232 mV |
| Higher Order | 72 mV | -- |
| Background (Offset) | 66 mV | 36mV |
| Mode Matching (%) | 77.4% | 93.9% |
In our typical fashion, I also measured the reflection scan of the cavity to work out or excess loss from the crystal and the escape efficiency at both red and green. No particularly important crystal position was chosen for this sweep.
| Measurement | Green REFL Volts | Red REFL Volts |
| Off Resonance | 820 mV | 6.00 V |
| On Resonance | 368 mV | 5.00 V |
| Background | 4 mV | ~0 |
| Escape Efficiency | 77.1% | 95.7% |
| Excess Crystal Loss | 0.46% | 0.5% |
This excess crystal loss for green is not corroborated by the green single-pass loss measurement, which suggests a green loss ~2x larger than what I measured with the green single-pass setup. However, as we noted before, this measurement technique relies on both the mirror reflection coefficients being known prior, and becomes particularly sensitive to variation in these mirrors as their added loss to the cavity becomes comparable to that of the crystal (i.e. near critical coupling). In this case, our predicted loss yields a predicted finesse that is much higher than what we actually see (in this case it predicts F = 238, we see closer to 140). So, I wanted to try an experiment where I fit the normalized reflection coefficient scan data (captured by the scope over time with "linear" PZT sweep) in such a way that the finesse is also taken into account. Taking this approach, we need no prior information about the mirrors to fit the reflection data. We only need prior information about the high-reflecting output coupler (R2 - to distinguish its effective loss from that of the rest of the cavity), the mode-matching coefficient, and to assume that the sweep is mostly linear (which is a stretch, but let's see how we do).
Essentially, scope data of both the reflection and transmitted light through the cavity is brought in to MATLAB, normalized and scaled correctly by the mode-matching coefficient, and fit against the theoretical Lorentzian line shape of the cavity. After normalization, the normalized power reflection coefficient can be expressed as PREFL = |(r1 - r2*beta*exp(i*phi))/(1 - r1*r2*beta*exp(i*phi))|2, where phi = 2*pi*t/TFSR, beta is the intrinsic field transmission coefficient through the rest of the cavity, and r1 and r2 are the field reflection coefficient magnitudes of M1 (the input coupler) and M2 (the output coupler) respectively. After scaling and fitting, the resulting fit against the reflection and transmission data from the scope for green is here. The fits on each resonance in the scan (raw here) give us estimated parameters for the reflectivity of R1, the intrinsic loss, the finesse, and the escape efficiency of the cavity. These are tabulated here for green assuming the output coupler (M2) has a 99.85% reflectivity and a mode matching of 77.4%.
| Intrinsic Loss (%) | R1 (%) | Finesse | n_esc | |
| Resonance 1 | 0.95% | 96.9% | 141 | 75.5% |
| Resonance 2 | 0.93% | 96.8% | 144 | 75.5% |
| Resonance 3 | 0.87% | 96.9% | 152 | 75.5% |
There is some variation in the fits from resonance to resonance. This is probably due to the limited memory depth of the scope (not enough data in the resonance to fit super accurately), small drifts over the sweep time, and the fact that the PZT actuation is not perfectly linear with applied voltage. Moreover, it seems like there is a strange thermal dragging/time delay between the transmission and reflection resonances. Nevertheless, the other two data points predict a loss much closer to what we measured in our single pass setup (~1%). As we stated before, it seems like this can be accounted for by M1 having a green reflectivity of ~97%, not 98%. To see if this is plausible, we will try to measure this mirror in situ the next time we take data on the cavity. For good measure, we also fit the reflection scans of the cavity at the seed wavelength (red) to compare with our typical escape efficiency calculation approach. The raw data and fits are attached here and here. The extrapolated parameters are listed below for red assuming the output coupler (M2) has a 99.85% reflectivity and a mode matching of 93.9%.
| Intrinsic Loss (%) | R1 (%) | Finesse | n_esc | |
| Resonance 1 | 0.56% | 86.7% | 42 | 95.6% |
| Resonance 2 | 0.48% | 88.3% | 48 | 95.5% |
Despite the slower scans taken here potentially worsening the drift-susceptibility of the data, the escape efficiency extrapolated from this data is fairly close to that of the simple dip-fraction calculation. In addition, the predicted reflectivity for M1 is on average about what we expect (~87.5%). However, the crystal loss is surprisingly high here. In the data that Maryrose measured last week, we were able to see much higher escape efficiencies for red than we are seeing now after re-installing the crystal. It is entirely possible that we are just currently measuring at a location in the crystal where there is some excess red loss (there are several cracks in this crystal due to the coating debacle), so we will try redoing the reflection scan over the crystal next time.
In any case, once we confirm that we have similar escape efficiencies as before, we should be getting pretty close to reinstalling the VOPO in the LASTI chamber and resuming squeezing experiments.