SQZ Site VOPO
- posted 21:56, Friday 12 September 2025 - last comment - 18:59, Tuesday 16 September 2025(12054)
PPKTP Crystal Characterization for LLO

[Daniel, Eric, Maryrose]

Initial Escape Efficiency Measurments for sPPKTP (#3) and PPKTP (circa 2018, #1 and #2):

We have been looking to test several crystals as possible candidates for the LLO site. We started (09/05) with testing out one of the sPPKTP crystals to confirm Daniel's measurements by measuring the escape efficiency and estimating the losses (table summary below). Later Eric provided some of the PPKTP crystals from the 2018 batch. They had previously been unopened, and one of them was exceptionally good (and was used in the H1 OPO we are planning to use as a replacement for the sites). We wanted to try out the rest of this batch in the LASTI setup to see if we could find another exceptionally good one. After testing two of them out in the LASTI cavity (numbered "1" and "2"), we found them to have nearly identical escape efficiencies and estimated losses. We proceeded to leave the "1" PPKTP in the LASTI and do further tests. A summary table of values is shown below for the measured mode matching percentage, escape efficiency, and estimated losses for red and green. Finesse measurements were also made for two of the crystal-cavity pairings.

Initial Loss Estimates and Escape Efficiency Measurements
Crystal Dip Fraction Mode Match % Mloss Mode Match with Loss est. Escape Efficiency % Finesse
sPPKTP, #3 0.9 91.5% 0.9975 0.901 97.4%  
  0.62 77.7% 0.99815 0.621   150
PPKTP, #2 0.93 96% 0.9989 0.931 98.3%  
  0.47 69.6% 0.9945 0.467    
PPKTP #1 0.96 93% 0.9988 0.931 98.2%  
  0.44 73.9% 0.9946 0.438   130

 

NLG Measurements with 2018 PPKTP Crystal #1:

We moved onto making NLG measurments across the cavity. These measurements showed that the PPKTP (#1) was pretty well aligned and had pretty uniform NLG across the crystal. Several scans were taken using estimated NLG measurements and NLG measurements with the cavity locked and optimizing the temperature and stage position. Finally, a rough scan near the center of the crystal was performed to obtain a range of acceptable temperatures for maximum NLG. Plots and tables are shown below for these measurements.

Initial NLG Rough Measurement Across Crystal
(T=32.23 C for all points)
Position (mm) Red w/ green Red no green NLG
-2 X X Edge of Crystal
-1.95 X X No useable co-res.
-1.45 -- -- Data deleted (~2.5)
-0.95 0.297 0.117 2.54
-0.453 0.299 0.113 2.65
0.04 0.253 0.111 2.28
0.537 0.253 0.111 2.279279279
1.028 0.259 0.111 2.333333333
1.521 0.243 0.111 2.189189189
2.019 0.257 0.111 2.315315315
2.505 0.245 0.107 2.289719626
2.786 X X Edge of Crystal

Average NLG from these measurements was around 2.36. Plot attached of this data.

 

Optimized NLG values for Position and Temp

Position (mm)

Temp (C) red w/ green red no green NLG    
-1.51 31.7 0.305 0.113 2.699115044
-0.5 31.88 0.309 0.113 2.734513274
0.483 31.83 0.257 0.113 2.274336283
1.483 31.89 0.281 0.115 2.443478261
2.464 31.8 0.269 0.113 2.380530973

Average NLG from these measurements was around 2.50. Plot attached of this data.

Finally, temperature vs NLG was measured. Scanning around -0.5 mm was done to find co-resonances at different temperatures. Plot of this data is also attached. We found the range of temperatures from 31-33 C to be a good setting for maximum NLG. Temperature controller was calibrated to the thermistor coefficients before this scan was done.

Discrepancies with Estimated Losses Between Escape Efficiencies, Finesse Measurements, and Daniel's Single Pass Crystal Tests:

We encountered a large discrepancy between the losses given by the escape efficiency measurements and those found by Daniel's single pass crystal measurements and the finesses we measured. We tried looking into this to see if we could nail down any inaccuracies in our measurments. We redid the escape efficiency measurment and found it to be in great agreement with the measurements taken earlier in the week.

Crystal Dip Fraction Mode Match % Mloss Mode Match with Loss est. Escape Efficiency % Finesse
PPKTP #1 0.96 94.7% 0.9989 0.932 98.3% 44
  0.42 73.9% 0.9942 0.423   126

To pin down what we think the green losses are in the previous batch of aLIGO crystals, we measured them in our single-pass setup from this alog. The full loss transmitting through the crystals (only #2 and #3 measured due to the fact that #1 is currently in the LASTI cavity) are listed here:

Crystal Trans PD Voltage (w/ Crystal) Trans PD Voltage (w/o Crystal) Loss
#2 0.459 V 0.468 V 1.92 %
#3 0.463 V 0.469 V 1.28 %

Using this as a baseline, we can predict what our green finesse should be given our mirror design (R1 = 0.98%, R2 = 0.9985%). F = -2π/ln(R1R2(1-Lxtal)). In the case where we believe we know all of our dominant loss terms (excluding possible excess loss from M3 and M4), we can compute a range of finesses in the LASTI VOPO to expect for all the aLIGO crystals in this batch (assuming they are similar). These are as follows:

Crystal Excess Loss (Lxtal) Finesse
#2 1.92 % 153
#3 1.28 % 182

We then measured the green finesse of the LASTI VOPO with crystal #1 installed. This turned out to be approximately ~120. Which is quite low and a far cry from the expected calculated values for the cavity finesse given our assumptions on the mirrors and measurements of the crystal loss out of the cavity (Maryrose mentioned that the finesse of crystal #2 and #1 looked very similar).

To further constrain the possible errors in this measurement, we made another green escape efficiency measurement of the VOPO with crystal #1 installed. In the process of this calculation, we estimate the total excess loss in one round trip through the cavity. Typically, we attribute this loss to the crystal. From this excercise we find that with our assumed mirror reflecitivites, we would expect the crystal's added loss to be on the order of 5000ppm, which does not agree with our single-pass measurement in the slightest.

Given that we know both our coupling condition and finesse to some degree from the reflection dip measurement associated with our escape efficiency and a simple cavity scan respectively, we hypothesized that the estimations of both the finesse based on measured single-pass crystal loss, and the excess loss from the crystal based on the cavity's green escape efficiencies may be flawed due to a flawed mirror. Next, we tested some different possible values for R1 to see if we could find a plausible problem with that mirror that might lead to this discrepency. If we suppose that M1's reflectivity was actually, 96.3%, we can recalculate the excess loss of the cavity (from the reflection dip) and the finesse estimate (with known crystal loss) again and find that the two estimates are closer to what we measured in the lab. In this case, our reflection dip measurement predicts the loss of the crystal to be closer to those measured directly in the single-pass setup. Moreover the additional coupling from the drop in R1 reduces the finesse to something closer to what we estimated with a cavity scan.

R1 R2 Calc XTAL Loss (Lxtal) Calc Finesse (Based on calc loss)
98% 99.85% 0.58% 228
96.5% 99.85% 1.2% 122
    Measured XTAL Loss Measured Finesse
    1.28-1.92% ~120

We also measured the reflection dip of the cavity @ 1064nm using the front seed path. Using this, we estimate the red escape efficiency, crystal loss, and finesse to be:

R1 R2 Escape Efficiency Calc  Red XTAL Loss (Lxtal) Calc Red Finesse (Based on calc loss)
87% 99.85% 98.1% ~800ppm 44.3

Our measured finesse at 1064nm is also around 44, so things make more sense with red than with green. When measuring the cavity with green light, it seems that we are in a regime where the reflection dip measurement is particularly sensitive to the reflectivity of the input coupler since it's coupling strength is smaller.

Images attached to this report
Comments related to this report
dmdesant - 16:01, Tuesday 16 September 2025 (12055)

Yesterday we pulled aLIGO crystal #1 back out of the LASTI VOPO to reinstall the SPPKTP crystal we (sort-of) like for more testing. In the process, we measured the loss of the aLIGO crystal on the single pass setup and found it to have a loss of around 1.8%. As far as loss goes, this crystal seems to be pretty good.

Unfortunately, in the process of reinstalling SPPKTP #3, I dropped the PEAK keeper for the crystal on the ground so Eric and I ran a batch of them (along with some other tools) through the air-bake oven. These are finished and we are now ready to continue with the reinstallation of the SPPKTP crystal.

dmdesant - 18:59, Tuesday 16 September 2025 (12056)

Note: Maryrose Barrios took a lot of this data and typed this nice summary up for us to have for our records. She and I just posted it with my alog account after we brainstormed a bit about our hypothesis that M1 may have a strange reflection coefficient.