Displaying report 1-1 of 1.
Reports until 16:48, Thursday 11 September 2025
SQZ Site VOPO
eoelker - posted 16:48, Thursday 11 September 2025 (12052)
O5 Oven assmblies soldered and glued

Today I soldered together the new oven assemblies for the SmarAct translation stages and glued the thermistors in place.  

For the soldering process, I did something losely based on E1700369, but modified John and Fabrice's procedure in a few ways.  The procedure:

  1. Class B clean all the required components (Peltier, thermistors, and copper oven componets.  Checking their ICS record, John and Fabrice air baked all components after ultrasonic cleaning.  DO NOT air bake the copper components.  This will cause the copper to become heavily oxidized and make it very difficult to adhere the InSn solder to the copper.  This may be why they needed to use HCl to get the indium solder to bond properly.  I air baked some of my copper wire a couple of months ago and could never get indium solder to stick to it after that.
  2. I used a hot plate covered in UHV foil like they did to heat up the copper components just enough to melt the InSn solder (melting point 118 C) but not high enough to melt the peltier solder (melting point ~160 C), placed the peltier on top, and then ramped down the temperature so the solder could set.
  3. I reversed the order of the soldering.  I attached the peltier to the top piece first (D1600067-v3) and then attached the combo to the base piece (D1600066-v4).  This allowed me to avoid using the soldering iron to attach the top piece.  Was a bit less cumbersome w/o having an iron involved.

I tugged on the components a bit and the assembly seems to be pretty strong.  I also glued in the thermistors to the top and reservoir; that process was straightforward.

Once the epoxy for the thermistors cures overnight, I will connectorize and then air bake the assembly for a few hours as a test to make sure it can withstand a thermal cycle to 100C.  We plan to ship the oven assemblies to Caltech and have the entire oven assembly vacuum baked at 100 C  (For the O3 oven build, we vacuum qualified the parts individually and the oven assembly was not re-baked after soldering and gluing the thermistors).  

 

Images attached to this report
Displaying report 1-1 of 1.