dmdesant - posted 17:46, Friday 08 August 2025 (12034)
Reprofiling LASTI HD Beams

Over the past few days, we have been reprofiling the beams and redoing the mode-matching setup in LASTI. We used the Phasics camera to perform all measurements

With phasics camera base at 19" from holes with screws holding the LO collimator, the LO beam parameters from the collimator are

  Major Minor Radial
M^2 (mm) 1.15 1.13 1.14
2W_0 (mm) 2.062 1.841 1.961
Z_0 (mm) -590.8 -983.5 -806.8

With phasics camera base at 7th hole back from far edge of table on SQZ path after first mirror, the SQZ beam parameters from the VOPO are

  Major Minor Radial
M^2 (mm) 1.12 1.13 1.13
2W_0 (mm) 0.164 0.170 0.170
Z_0 (mm) 203.6 196.8 200.3

Relative to HD BS, LO collimator screws are 29" away (-z direction)

Relative to HD BS, back of table holes are 28" away (-z direction)

Detector plane of the camera is 38.5mm forward of the mounting hole position (-z direction)

Relative to the HD BS, the LO waist positions are -(29inches-19inches)*25.4mm/inch - 38.5mm offset from measurement (-292.5mm)

  Major Minor Radial
M^2 (mm) 1.15 1.13 1.14
2W_0 (mm) 2.062 1.841 1.961
Z_0 (mm) -883.3 -1276 -1099.3

Relative to the HD BS, the SQZ waist positions are -(28inches-6inches)*25.4mm/inch - 38.5mm offset from measurement (-597.3mm)

  Major Minor Radial
M^2 (mm) 1.12 1.13 1.13
2W_0 (mm) 0.164 0.170 0.170
Z_0 (mm) -393.7 -400.5 -397

Using these beam parameters, a new mode-matching solution was found with one less lens in the SQZ path and one more lens in the LO path.

LO Lens Z Position (Relative to HD BS in meters) ROC (m)
1 -0.6582 0.100
2 -0.3908 0.050
3 -0.1202 0.150

 

SQZ Lens Z Position (Relative to HD BS in meters) ROC (m)
1 -0.6273 0.100
2 -0.1201 0.100

The theoretical beam propagation model is as follows.

After correcting mode matching, we measured the following visibility

  PD1 PD2
DC Offset (mV) 5.6 -5.6
Bright Fringe (mV) 2080 2080
Dark Fringe (mV) 25.6 14.0
Visibility 98.1% 98.1%

In theory, we may now be able to observe 10dB of squeezing with this readout. Next steps include recovering our last-known good crystal configuration and measuring squeezing again as a function of NLG