I reallized on July 29th that once again, the Meerstetter TEC-1163 stopped measuring a resonable resistance through its HR-1 channel. (This device is the temperature controller for the prestabilization setup in the vacuum tank.) I tried using a different pin pair in X-14 (this worked the last time this device had issues 11938), as well as the usual resetting things in different order. The resistance from the in-chamber NTC sensor was around 12 kΩ, which is very resonable, although the value read thru the 23-bit ADC was MΩ. Magically, the device now works. The only thing I can say that helped it start working is replacing the NTC sensor pins with the two legs of a standard 13 kΩ resistor, at which point the device started reading reasoable resistance values again. After this, I plugged the NTC pins back in and suddenly it started working. I had reset the 23-bit ADC through the software before, with the and without the resisitor attached, without any success.
Some notes: