Today I tried to do some loss hunting with the spare VOPO. As I diagnosed yesterday, M1 appeared to be scattering some excess light. I tried drag wiping the optic in-situ, but wasn't able to make much progress on the escape efficiency.
I then attempted to measure the escape efficiency without the crystal by moving the crystal out of the beampath using the translation stage. I wanted to see whether the excess loss was coming from the crystal or the mirrors. This misaligns the cavity, of course, so I tried to recover alignment by moving M4 a bit. Unfortunately I wasn't able to align to a reasonable mode and I suspect that the cavity is unstable. According to the design document, it should be possible to form a stable cavity w/o the crystal and Masayuki was able to do it at LLO it in 2020. However, the design document says that the cavity is quite sensitive to the M3-M4 distance. I tried varying this distance a bit, but gave up after an hour or so.
Finally, I decided to swap out M1 with the second M1 assembly. I was then able to inspect the old M1 mirror more carefully using a green flashlight. Sure enough, there does appear to be a very faint cloudy spot in the center of the optic that doesn't come off with drag wiping. Tomorrow I will inspect with Slawek's microscope and may try First Contact.
Finally I realigned the cavity with the new M1 and measured the escape efficiency. I was able to get the coupling for the front seed path up to 97%. Now I'm measuring a reflection dip of 88.5%, which corresponds to an escape efficiency is up to 96.9%. Together with M2 (1000 ppm transmission) and M3/M4 (70 ppm transmission each), this corresponds to an excess loss of around 1250 ppm. This would be well within spec for the AR coating on this batch of crystals, though the previous batch of crystals were further below the spec value (Masayuki's measurements suggests that they had < 360 ppm per surface).
I then locked the cavity and took some photos. The green scattering looks much improved. However, I noticed that the beam is no longer centered vertically in the crystal after realigning the cavity. I will attempt to raise the oven tomorrow to center the beam. Hopefully, the AR coating is under spec and I'll be able to recover more of the missing efficiency by centering the beam.
Also, while I was aligning the cavity, one of the peltier leads broke off. We will repair this tomorrow.
This afternoon, i loosened the screws attaching the oven mounting brackets to the oven baseplate and adjusted three point support to raise the crystal slightly. I centered the 532 nm beam on the crystal and am still measuring an 88.5% reflection on resonance with the same mode matching as yesterday, so the escape efficieny is unchanged.
Next things to try:
If any of the other crystals seem promising, I might try swapping the crystal at some point to see if I can boost the escape efficiency further.
Slawek managed to clean the M1 mirror we removed with some more forceful "drag" wiping. We'll try this mirror again at some pont with the second spare.