SQZ General
dmdesant - posted 18:36, Friday 25 July 2025 - last comment - 16:09, Thursday 14 August 2025(12019)
LASTI VOPO Escape Efficiency Remeasurement

Today, I remeasured the escpape efficiency. To do so, I assembled a more permanenant back seed path which is similar to the rest of the fiber paths into the cavity. After alignment, I checked the mode matching with a cavity scan and noted the peak transmission values for the TEM00 mode and all higher order modes visible on the CLF PD DC monitor voltage

Mode Overlap Measurement (Transmission Scan)

Mode Detector Volts
TEM00 0.3900
TEM02/20 .07425
HOM1 .02775
HOM2 .02445
Detector Baseline .02100

This corresponds to a mode matching (overlap) of about 85.3%. Then looking at the reflection dip for the TEM00 mode on the reflection scan and following the same approach we did before in this alog, we collect the following data

Reflection Scan

Measurement Detector Volts
Peak Reflection (Off Res) 1.369
TEM00 (On Res) 1.264
Detector Baseline .00093

This measurement of the reflection dip corresponds to a calculated escape efficiency of 97.6%, which is a bit lower than we had before and could be due to the fact that the reflection PD was clipped in that measurement. Moreover, we did not note where on the crystal we had taken the previous escape efficiency measurement. This measurement was taken with a crystal temperature of 32.1C, and a stage position of -0.7695mm. With this back seed path in permenantly now, it should be relatively easy to repeat this measurement a few times in different spots on the crystal as we continue to optimize.

Comments related to this report
dmdesant - 16:09, Thursday 14 August 2025 (12041)

Note: There is an error here... back/rear actually should be front...