Today, I remeasured the escpape efficiency. To do so, I assembled a more permanenant back seed path which is similar to the rest of the fiber paths into the cavity. After alignment, I checked the mode matching with a cavity scan and noted the peak transmission values for the TEM00 mode and all higher order modes visible on the CLF PD DC monitor voltage
Mode Overlap Measurement (Transmission Scan)
| Mode | Detector Volts |
| TEM00 | 0.3900 |
| TEM02/20 | .07425 |
| HOM1 | .02775 |
| HOM2 | .02445 |
| Detector Baseline | .02100 |
This corresponds to a mode matching (overlap) of about 85.3%. Then looking at the reflection dip for the TEM00 mode on the reflection scan and following the same approach we did before in this alog, we collect the following data
Reflection Scan
| Measurement | Detector Volts |
| Peak Reflection (Off Res) | 1.369 |
| TEM00 (On Res) | 1.264 |
| Detector Baseline | .00093 |
This measurement of the reflection dip corresponds to a calculated escape efficiency of 97.6%, which is a bit lower than we had before and could be due to the fact that the reflection PD was clipped in that measurement. Moreover, we did not note where on the crystal we had taken the previous escape efficiency measurement. This measurement was taken with a crystal temperature of 32.1C, and a stage position of -0.7695mm. With this back seed path in permenantly now, it should be relatively easy to repeat this measurement a few times in different spots on the crystal as we continue to optimize.
Note: There is an error here... back/rear actually should be front...