Today I realigned our homodyne to 96.5% visibility before characterizing our squeezing as a function of NLG. It seems like something has shifted from previous alignment attempts or I was not getting as much visibility as I had thought initially.
I also roughly calibrated the measured power on the pump reflection beam coming out of the chamber relative to the pump input power off resonance. With 26mW pump incident on the OPO, we have ~24.6 mW measured on refl beam outside the chamber.
Then, by adjusting this power to a few points over a wide range, I quickly measured the roughly optimal NLG at a given pump power and then locked everything and measured the squeezing and anti-squeezing levels on the homodyne. This data is shown below.
| Pump REFL (mW) | Temp (C) | Stage (mm) | NLG | Squeezing (dB) | Anti-Squeezing (dB) |
| 7.0 | 32.45 | -0.7695 | 3.36 | 5.5 | 8 |
| 12.2 | 32.36 | -0.7695 | 6.26 | 7 | 11.5 |
| 18.0 | 32.24 | -0.7695 | 12.8 | 7.8 | 15.3 |
| 25.0 | 32.10 | -0.7695 | 39.3 | 8.0 | 20.3 |
I also noticed that having the SR785 on DC coupling greatly reduced its dynamic range and its own noise floor was beginning to impact our measurements. This was fixed pretty easily by just AC coupling the input, which yielded much more predictable and consistent data over a wide set of input attenuator settings and shouldn't matter at the frequencies I am looking at here anyway. These squeezing levels are an "eyeball" average over the spectrum the analyzer was displaying, but I would say that they are a low conservative estimate of the real average above 400 Hz. Once this data is fit to the theoretical curves for this, we will begin hunting for losses and additional noise that could be impacting our ability to reach 10dB of squeezing.
Today I post-processed this data and got an estimate for the total detection efficiency and phase noise, along with a sanity check for the threshold power of our OPO at the moment. Attached are plots with fits for the nonlinear gain of this cavity as well as the squeezing and anti-squeezing levels we saw yesterday.
The results of these fits predict our threshold power to be around 36 mW with 95% confidence bounds (35.17 mW, 36.18 mW). The squeezing fit predicts the total detection efficiency to be around 85% with 95% confidence bounds (81.6%, 88.2%). The phase noise is so small that this fit is insufficient for estimating phase noise (as there is no inflection point in the experimental data). If we trust this characterization, the current 85% detection efficiency of this setup limits us to a maximum theoretical measured squeezing of 8.2 dB, so now we may be up against the loss limits of our detection system. 85% detection efficiency is also fairly believable given some ballpark estimates and loss measurements we have already made.
At the moment, our estimated losses are as follows
Given that we need our detection efficiency to be clear over 90% in order to see 10dB of squeezing, we stand to gain the most efficiency in reoptimzing our homodyne detector, which currently contains more lenses in the sqz path than are strictly necessary. Moreover, if the seemingly low homodyne visibility we were seeing yesterday were improved to 98% or better, it would push this current estimate up to around 91.4%, at which point seeing 10dB could be possible.
One final note is that our escape efficiency measurement from last week may be slightly off as Eric noticed that his escpae efficiency measurement on the spare site VOPO was affected by the reflection of the rear seed path being clipped by the PD he was using at the time. We has a similar problem on the LASTI cavity, so will be remeasuring the escape efficiency again to pin this down as accurately as possible.