Figured out that my issue yesterday was clipping on the refl PD for 1064
Now I'm getting something that makes more sense for the escape efficiency
mode mismatch was constant at 100 mV, so the change is in TEM01
Unfortunately our escape efficiency sucks right now. Values here assume an M1 reflectivity of 93% and an M2 reflectivity of 99.9%
| Position | Reflection off of Resonance (V) | Reflection Dip (V) | trans TEM00 (V) | trans HOM Sum (V) | Excess Loss | Escape efficiency |
| -1100 | 5.28 | 0.880 | 2.24 | 0.220 | 0.27% | 95.4% |
| -2000 | 5.32 | 0.880 | 2.20 | 0.220 | 0.27% | 95.4% |
| -2900 | 5.32 | 0.880 | 2.20 | 0.220 | 0.27% | 95.4% |
| -3800 | 5.32 | 0.880 | 2.20 | 0.220 | 0.27% | 95.4% |
| -4700 | 5.32 | 0.840 | 2.12 | 0.310 | 0.26% | 95.5% |
| -5800 | 5.32 | 0.840 | 2.12 | 0.310 | 0.26% | 95.5% |
| -6900 | 5.32 | 0.840 | 2.12 | 0.310 | 0.26% | 95.5% |
| -8000 | 5.32 | 0.840 | 2.15 | 0.280 | 0.26% | 95.5% |
| -9000 | 5.32 | 0.840 | 2.14 | 0.290 | 0.26% | 95.5% |
I'll note that we see the same HOM spacing (0.27 FSR) and change in alignment across the wedge (~3%) as in the 532 nm scan. I estimate that the finesse is roughly 86