Did a reflection dip measurement to determine the 532 nm losses. Solving for the excess loss actually gives a double valued result, either an excess loss of 5% or an excess loss of 0.8% assuming an input mirror reflectivity of 98%. However, the latter value is more consistant with the measured finesse for 532 nm (measured 190, calculate 220 with 0.8% and 90 with 5% excess loss). I could constrain by using the transmitted power value as well, but haven't calibrated these voltages in terms of power yet.
This is quite good compared to some values observed towards the end of O3 (was 2-3% with an older crystal): https://dcc.ligo.org/LIGO-T2000150
Presumably this is just because this is a brand new crystal?
Again we see about a 3% change in misalignment across the entire wedge.
I tried to measure escape efficiency at 1064 nm as well, but things aren't making sense yet. Need to debug tomorrow.
| Position | Off Res | On Res |
TEM00 trans |
TEM01 Trans |
02+20 trans |
Excess Loss |
| -200 | ||||||
| -1100 | .222 | .094 | 1.86 | .140 | .460 | 0.8% |
| -2000 | .224 | .092 | 1.92 | .120 | .440 | 0.8% |
| -2900 | .226 | .090 | 1.86 | .120 | .460 | 0.8% |
| -3800 | .226 | .090 | 1.94 | .60 | .460 | 0.7% |
| -4700 | .226 | .090 | 1.98 | .60 | .440 | 0.7% |
| -5700 | .226 | .090 | 1.98 | .60 | .440 | 0.7% |
| -6800 | .226 | .090 | 1.96 | .100 | .420 | 0.8% |
| -7900 | .226 | .090 | 1.96 | .160 | .420 | 0.8% |
| -9000 | .228 | .090 | 1.98 | .160 | .420 | 0.8% |