SQZ Site VOPO
eoelker - posted 10:59, Tuesday 22 July 2025 (12008)
HOM Scan and PZT resonances for H1 Spare VOPO (SN3)

PZT scan reveals the following HOM spacing for 532 nm

HOM Delay from TEM00 Fraction of FSR
TEM00 0 0
TEM40 17 ms 1.08
TEM10 58 ms 0.27
TEM20+02 116 ms 0.54
TEM00 215 ms 1.0

This should be OK.  The first 10 modes are several FWHM away from the TEM00 resonance.

The PZT resonances are a bit lower than some that Maggie measured for O3:  4.8 kHz for M4 and 5.6 kHz for M2 ( maggie was getting more like 6-7 kHz).  They seem to be OK, however.  I can still lock with a UGF > 1kHz for the OPO with an appropriate notch filter.