Reports until 10:50, Tuesday 22 July 2025
SQZ Site VOPO
eoelker - posted 10:50, Tuesday 22 July 2025 (12007)
NLG Scan across 10 co-resonances for H1 Spare VOPO (SN3)

This morning I did a scan across all the co-resonance positions and characterized the NLG at each for the phase matching temp I measured yesterday 35.25C.  I believe that I was at point -1800 yesterday

Green power incident 2.2mW and was fairly constant across the measurement. 

Peak power off co-resonance for 1064 nm = 320 mV

Stage Position Delta V NLG
-200 1064 losses high  
-1100 2.58V 8.1
-1800 2.56V 8.0
-2700 2.44V 7.6
-3600 2.40V 7.6
-4650 2.44V 7.6
-5800 2.64V 8.3
-6900 2.80V 8.8
-8000 2.54V 8.0
-9000 2.58V 8.1

Crystal edges (where cavity flashes go away) are at 0 and -9700

 

NLG looks fairly uniform.  I did not optimize the phase matching temp at each position, i just used 35.25C for all.