eoelker - posted 19:00, Monday 21 July 2025 (12006)
Co-resonance scan of the H1 VOPO (SN3)

I did a scan on Friday to find all the co-resonance positions on the crystal when the temp was set to 35.25C.  The edges of the crystal were at 0 and -9700 counts on the translation stage software (it does not keep track, so I did this manually).  I found 10 coresonances in all, though the one at -200 had very poor transmission for 1064 nm and is likely not useable.  I was curious to see how much the 532 nm cavity misalignment changes across the entire crystal.  I started w a slight misalignment (a couple percent) so i could keep track of the mode.  I saw values between 2.3% - 4.7% across the entire face of the crystal.  The data was a bit noisy so i suspect some of this is just drift from the temp fluctuations/air currents in the cleanroom or from me bumping the mirrors when I was touching up the fiber polarization.  I haven't found a super precise record of this anywhere, but this doesn't seem like a huge level of change over the entire wedge.  Its not obvious to me that there is even a crystal alignment that would get rid of this effect entirely.  I'll have to look back at John's design document for the VOPO to see if he estimated the magnitude of this effect.  

All values here are for 532 nm.  I'll do this for 1064 when I'm characterizing the escape efficiency tomorrow.  

Stage pos TEM00 TEM01 TEM02+20 % misalignment
-200 1.00 0.040 0.240 3.13%
-1100 1.03 0.050 0.250 3.75%
-2000 1.10 0.040 0.260 2.86%
-2900 1.10 0.032 0.260 2.30%
-3800 1.10 0.056 0.260 3.95%
-4750 1.14 0.056 0.270 3.80%
-5700 1.15 0.064 0.270 4.30%
-6800 1.18 0.072 0.280 4.70%
-7900 1.18 0.064 0.280 4.2%
-9000 1.18 0.064 0.280 4.2%

 

Displaying report 1-1 of 1.