dennis.martynov - posted 05:56, Wednesday 18 June 2025 (11986)
T240 X/Z diagonalisation
Rich and I had a peculiar discussion about the precision of the T240's transformation from its Galperin configuration to XYZ. While I think it is straightforward to diagonalise Z with a sub-percent precision (because it should not see any ground tilt), it might be harder to achieve clean X and Y signals. In log 11851, I witnessed a larger T240 signal compared to GS-13 and C-6D at microseism and became suspicious that T240's Z->X coupling is around 5%.

Since it might have been only a miscalibration at that time, I have tested T240's Z -> X coupling by stabilising ISI with GS-13 in Z (without sensor correction). All other degrees of freedom were stabilised with C-6D (with sensor correction for X and Y). In this configuration, we have a significant ISI motion in Z around microseism (limited by the ground motion) but smaller motion in horizontal degrees of freedom.

The attached figure shows the result. It turns out that the T240's transformation is better than I expected before, and Z -> X coupling is smaller than 1 / 250: orange vs red at 250 mHz. The coupling is likely even smaller because it seems that T240 sits on some noise: either on its own noise or on the C-6D's laser frequency noise (since we do not run LPS frequency stabilisation servo).