We continued coupling studies with the fused silica cavity and prisms, this time using 2.5 µm aluminum foil applied using the procedures described previously.
Swadha has set up the cavity in the vacuum tank for signal injection and extraction.
The following describes the optical configuration when the cavity is resonating, with powers given by a power meter. The laser is an NPRO (1064 nm) and the input polarization is p.
- On the input to the cavity, 17.0(5) mW incident on the AR surface to the coupling prism.
- At least two ghost beams are visible on reflection from the input side of the prism. Their power is 3.5(5) mW. They are oriented "outside" the incident beam, i.e., away from the prism apex. This is consistent with the 47° opening angle of the prisms.
- On the reflection from the cavity, 12.4(5) mW is reflected total.
- The prisms are AR coated for 1550 nm, so I suppose we should not be alarmed by the ~15% AR reflections that we infer from the above.
- There is a main beam on reflection and one ghost beam. The ghost beam is on the "inside" relative to the main reflected beam (toward the prism apex). Its power is 0.5(5) mW.
- The transmitted power is 2.8(5) mW. There is only one transmitted beam visible on the card. The beam intensity pattern is mostly round, i.e., nominally TEM(0,0). The TEM(0,22) nearly coresonant mode is also visible.
- Downstream on transmission, there is a 50/50 beamsplitter, with one output port going to a PDA100A. The transmitted voltage here is 450 mV on the 0 dB setting. The responsivity at 1064 nm is 0.27 A/W. The transimpedance is 1.5 V/mA. One would therefore infer 1.1 mW on the photodiode, or 2.2 mW before the transmission beamsplitter.
- On the fourth, nominally unused, port, a round beam is visible. I did not measure its power, but it could plausibly be a ghost beam.
- From the above numbers, one infers (assuming 15% AR reflection here) there is 14.5 mW incident on the input FTIR interface, and 3.3 mW exiting the output FTIR interface.
The cavity finesse in p is something like 700, and the finesse in s is higher. I did not make a more detailed measurement yet than eyeballing on the oscilloscope.
Now, some comments about mode matching and HOMs:
- We did not attempt a careful modematching setup. Nonetheless, the modematching as inferred from the TEM02/TEM20 mode height is quite good. With 324 mV for TEM00 on the PDA100A, we then find 13 mV for TEM02 and TEM20.
- Referring to the attached finesse simulation, which simply uses the
knm library to compute overlaps into the cavity's TEM00, TEM02, and TEM20 modes from a mismatched Gaussian mode at the input, we should infer that we are coupling 90% of the input light into the cavity, assuming the beam is reasonably clean.
- Throughout all of our studies with this cavity, the transmitted power of the nearly coresonant TEM(0,22) mode has been about 50% the height of the TEM(0,0) mode. This was true with the Planex laser and also with the 1.8 µm foil. It is also true (at least by eye) when resonating s instead of p. The mode height is also much greater than any other HOM. A potentially troubling inference to make is that the cavity internally couples these two modes. Taking 2.0 MHz as the previously ascertained splitting between these two modes, and assuming a finesse in p of 700, this would imply that on resonance for TEM(0,0), the power in the TEM(0,22) mode is about 10% of the TEM(0,0) power.
I turned the input power down to 1.84(5) mW, giving 1.32(5) mW in reflection. I set up an NF1811 in reflection and found the following:
- I measured 1.17(5) mW incident off resonance (again these powers are with the power meter).
- The voltage from the dc monitor is 1.10 V, which is not consistent with the incident power assuming a responsivity 0.7 A/W and a transimpedance of 10 V/mA. We saw this previously with this NF1811 using the Planex during the 1.8 µm tests. The beam seems reasonably small on the photodetector (the width in pitch seems smaller than the active diameter, and width in yaw seems about as small or smaller than the active diameter).
- In any case, the voltage on resonance was 0.56 V. Since the mode matching inferred from the HOM spectrum on transmission is high (greater than 90%), we might infer that the ~50% visibility on reflection is due to under- or overcoupling of the cavity.