hudsonl - posted 17:22, Wednesday 28 May 2025 (11972)
Phasics Camera
[Hudson, Jacques]

We tested out the Phasics camera for mode matching the beam out of the Mach Zehnder interferometer with the breadboard OPO homodyne's local oscillator. After about 6 hours of time in the lab, including time spent figuring out how to use the Phasics camera, we achieved 97.1% mode overlap (visibility). Profiling the final beams with the phasics camera, the computed local oscillator and MZI beams have a computed overlap of 96.9%, consistent with our measurement on the homodyne itself. 

As a reference point, I spent ~20 hours in the lab the previous week attempting to mode match these two paths using a CCD beam profiler and could only manage to achieve 93% visibility.

Some observations:
1) If the beam is too small or too large on the camera, the computed waist size and position will be inaccurate. However, the computed numbers do not include error bars, so it can be somewhat hard to tell when this is the case. The sweet spot seems to be a beam diameter of 1 mm or so.
2) A good work flow seems to be
     i) Take a beam profile of the seed and target beams with the Phasics camera at a fixed location
     ii) Use some mode matching software such as a la mode determine the mode matching lens locations
     iii) Place both lenses and use the Phasics camera to coarsely position the lenses in the viscidity of the locations computed with the mode matching software.
     iv) Fine-tune the lens positions using a homodyne diode.