- posted 13:22, Monday 19 May 2025 (11963)
BHQS Prototype Center of Mass tests

JD, Regina

COM Tests

We are trying to verify the center of mass of the top and UIM masses in an effort to place the wire break off points as close to d=0 as possible. We set up a single mass suspended by 2 wires and placed 2 displacement sensors on either end of the top of the mass (see figure 1). The entry on the old ilog describes the displacement sensor setup. In order to find the d-value, we measured the pitch motion and calculated the vertical offset based off the resonance. We kept the length of the wire the same.

We found that we need a resolution of 2 mHz to get to our target of measuring the d offset to 200 microns. We initially set the bandwidth to 512 s (0.002 Hz), but will probably do our next tests with 1024 s. We conducted experiments by changing the d-value of the wire by adding precision shims in between the clamp and the mass. Fig 3 shows the experimental results, and Fig 2 shows the modeled frequencies that we expect by fitting the values to match the longitudinal mode (higher frequency peak).

We had an idea of where the breakoff point should be compared to where the center of mass is, but we weren't not certain. We expected the d-value to be around 18mm above the COM, and we changed the value around to fit the resonances in the power spectrum. For this first run through, we didn't put the flags and balance masses on the mass which meant the COM is lower than what we expect it to be in the full setup. For the next setup, we'll try to recreate the full setup in more detail.  

Cage

The parts for our cage around the masses came, and we began assembling it in the high bay. These parts allow us to align the masses precisely to one another and to the sensors. They also have space for us to add stoppers to hold the masses. Fig 4 shows the whole cage (we didn't put the brackets around the UIM yet). fig 5 shows a close up of one of the BOSEM holders we are testing out that has a kinematic coupling attachment. 

 

Images attached to this report