Reports until 18:51, Wednesday 14 May 2025
Heisenberg IFO General
hudsonl - posted 18:51, Wednesday 14 May 2025 (11960)
MZI Locking
I switched the MZI lock over from using an LB1005 to a Red Pitaya. We decided to make this switch since locking the MZI exactly at mid-fringe seems critical to saturating the waveform quantum Cramer-Rao bound, and a computer is better suited to this task than a person looking at a scope. The fundamental issue is that the MZI error signal is small, and there are DC offsets on the differential receiver's output, the LB1005, amplifiers, etc.

The feedback loop is now differential receiver -> SR560 (50x non-inverting gain) -> Red Pitaya -> PZT amplifier -> MZI PZT

In contrast to the setup's previous configuration, there is no analog amplifier between the red pitaya's output and the PZT amp. Similarly, we added the SR560. Both of these were done to minimize effects of digitization in the RP's inputs and outputs. The red pitaya curve fits the differential receiver's output vs PZT drive signal to determine the setpoint corresponding to mid-fringe (this assumes the signals on the differential receiver are balanced, which should be verified before running the experiment). This is then the setpoint of the PID loop. 

We can achieve lock bandwidths of up to 5-10 kHz, but want to restrict the bandwidth to be less than 1 kHz so the 8 kHz modulation tone is much greater than the UGF and is not suppressed. From the transfer function, we see that the PZT mirror assembly has a structural resonance at about 700 Hz and a PZT resonance at about 25 kHz.

The script seems to acquire lock consistently and stays locked for >15 minutes.