Reports until 12:17, Wednesday 07 May 2025
TIR General
Swadha - posted 12:17, Wednesday 07 May 2025 - last comment - 15:39, Friday 09 May 2025(11953)
Switching to higher finesse cavity

Evan, Swadha

We had previously coupled the input and output prisms to the cavity using copper foil of estimated thickness ~1um, which gave us a cavity finesse of 76. We now switched to Aluminium foil of estimated thickness ~1.8um. We expected lower transmission coupling and thus higher finesse, which we achieved.

Switching the output coupling:
We started by switching the output prism coupling to Aluminium, and unscrewing the input prism to increase gap size for higher finesse. We re-aligned slightly to recover the TEM00 mode. Measured data:
1. In p-polarization: finesse = 1064
2. In s-polarization: finesse = 3500
However, with the input prism unscrewed, the setup was no longer stable and very sensitive to vibrations.

Switching the input coupling:
We then also changed the input prism coupling to Aluminium, and re-aligned. Measured data:
1. In p-polarization: finesse = 833 (TEM00 mode here, mode structure here).
2. In s-polarization: finesse = 1820 (TEM00 mode here)
3. Cavity p-s splitting = 73.4 MHz (data here)

We were able to lock the cavity in both polarizations.

Non-image files attached to this report
Comments related to this report
evanhall - 13:27, Wednesday 07 May 2025 (11954)

Attachments show expected cavity coupling properties assuming 1 nm rms microroughness on the internal faces, a bulk absorption of 5 ppm/cm, and equal input and output transmission values. This is the basis on which we infer the copper leaf was slightly more than 1.0 µm in situ (previously, we had estimated 0.6 µm by folding a square of leaf four times and measuring with a micrometer). It is the basis on which we infer the 1.8 µm aluminum microphone foil is perhaps more like 2.2 µm in situ. It could be creasing, although visually the foil looks quite flat when viewing the interfaces.

We had to crank somewhat hard on the front setscrew to achieve what we think is something close to critical coupling in p (it seemed to be very undercoupled before). The mode matching is still quite poor, so inferring anything quantitative about mode matching right now is difficult.

Non-image files attached to this comment
evanhall - 15:39, Friday 09 May 2025 (11957)

The attached sympy script computes the transverse mode spacings (92 MHz tangential, 65 MHz sagittal), and the expected waist radii (0.41 mm tangential, 0.48 mm sagittal at 1064 nm), assuming nominal cavity parameters (side length 5.08 cm, ROCs both 5.0 m). I think we missed a factor of 2 earlier and the FSR is in fact 1.44 GHz. The previously stated finesse values need to be divided by 2. In that case, the inferred leaf thicknesses are something like 0.7–0.8 µm for copper and 2.0 µm for the microphone foil — so some 0.2 µm of excess or so (perhaps still creasing, but less than we thought).

Looking at the camera, the nearly co-resonant mode that we observe is some TEM(0,n) with n of order 20. Indeed, taking the FSR divided by the tangential TMS indicates we should expect the TEM(0,22) mode about 1.3 MHz below the 00 mode. Instead, we see the co-resonant mode at about 2.0 MHz below, which could be consistent with the TEM(0,22) if the curved faces are about 0.1% more planar than 5.0 m.

I am also attaching an archive of the raw oscilloscope data for the above plots. The file numbering is

  • 01: sweep across TEM00 s mode
  • 02: sweep across TEM00 p mode
  • 03: sweep from TEM00 to TEM02 mode in p, with power pumped into first-order modes via misalignment
  • 04: sweep from TEM00 mode to TEM20 mode in p, with power in first-order modes minimized (can be used to infer something about the current mode mismatch)

The calibration was done by driving the EOM while sweeping the laser current. A 49.2 MHz drive produces sidebands in transmission corresponding to 1.44 V of input to the laser driver, hence 34 MHz/V. The thermistor readback was 7.27 kΩ, and the current knobs were 6′00″ coarse and 7′30″ fine.

Non-image files attached to this comment