I fit four oscilloscope time series measurement of the Teraxion 1550 nm laser in an asymmetric Michelson (arm length difference 17.5 cm) modulated with a sine wave (f = 1 kHz, A = 1V). The fits and corresponding residuals are shown in the first four figures. Fit parameters up to the fourth harmonic are shown in the table in the last file. m_i are effective modulation depths defined as m_i = 2*pi*(\Delta f)*\tau, where \Delta f is frequency modulation depth, and \tau is the optical pathlength delay, and m1 corresponds to the carrier frequency. The number of significant digits for the fit parameters corresponds to the fit errors from the covariance matrix (curve_fit). All measurements were taken with the same modulation, so parameter averages of the four measurements are given in the table and below. m1 = 2.81 +- 0.09 m2 = 0 m3 = 0.004 +- 0.008 m4 = 0.006 +- 0.007 If needed, a more precise linearity measurement can be made with cymac.
Raw data from RIGOL oscilloscope and Python script that generates the plots are attached below.
We noticed that the oscilloscope's ADC quantization steps are large for these measurements, so smoothing the data gives better fit results and more information. Attached are updated plots, fit parameters up to m6, and Python script. New modulation depths: m5 = 0.005 +- 0.004 m6 = 0
The attached script fits the nonlinearity using MCMC instead of least squares. It is slower and requires some babysitting of the sampler burn-in, but should explore the whole parameter space in the case that least squares might fail.
Neither of our scripts yet account for the modulation-induced AM, which should be added before making firm conclusions about the nonlinearity of the FM drive.