begum - posted 18:11, Tuesday 22 April 2025 (11938)
TEC recovery for prestabilization in-chamber temperature

After the power outage, Meerstetter TEC-1163 HR-2 (high resolution) input stopped reading resistance, so I was unable to recover the temperature control loop with the previous configuration. It seems this is an issue about the HR-2 input specifically, which coincides with the power outage for some reason. Low Res input 3 shows -90 °C all the time, although there is nothing connected to it. This threw me off for some time.
I changed the input to Connector 14 [attached photo top left connector], Pins 1&2 (instead of 3&4) and the sensor output (or HR input) is online on the software again. In case we need to recover this again, screenshots of the main screen, the HR1 input settings, TEC channel settings and PID controller screen are attached. This device is supplied with 32 V bias.
Note that the cables coming from the sensor are not directly soldered onto the pins but rather soldered in crimp terminals, so a firm pull will dislodge them, beware. 

Thanks to Eric for remote support from Vienna.