Sasha, Evan We fixed the grounding issue in RetroSense v3 and there is now no current leakage. The solution is using plastic screws on the optical head for isolation, but for future designs it would be preferable to use non-conducting components in the parts surrounding the LED. I centered the sensor and chose the distance between the LED and the retroreflector such that the diagonal signal in the centered position is equal to zero. The most recent noise measurement with this configuration is shown in the first figure. It includes the measurement of the SR560 preamp noise, since our dark noise is now at picometer level at frequencies above 1 Hz. The second figure shows the RetroSense sensitivity, range, and how voltage from different channels depends on displacement along the x-axis. Linear range is about [-0.2 mm, 0.2 mm]. This RetroSense version could already be placed on a suspension prototype for testing, so I will be looking into connecting it to the required readout. Measurement details: dV/dx = 51.6 V/mm V_sum = 14.0 V I_sum = 549 uA Shot noise = 8.36 pm/rtHz