Displaying report 1-1 of 1.
Reports until 13:27, Thursday 17 April 2025
SQZ General
wenxuan - posted 13:27, Thursday 17 April 2025 (11934)
OPO loop recovered after broken PZT chassis

Daniel, Wen

We noticed the PZT driver chassis was no longer working when we tried to sweep OPO M4 PZT. All other driver ports of the same chassis were also dead. We switched to the chassis above (2nd from top) to recover the OPO loop. 

The loop was able to lock, but the loop stability was noticably worse than before. The OLTF looks fine but the gain setting has to be reduced 10 dB. The phase margin of the new loop was similar with before. We also checked the PDH error signals and phasing, which also looked fine. After some searching, we found the new PZT chassis is HV version and clips at high voltage. We then switched to the 1st chassis from top and solved the problem. The OPO CMB gain is now 12 dB with UGF 1.3 kHz and 60 deg phase margin (as seen in the plot).

The green in-vacuum fiber collimator was shaky as mentioned in previous logs. The problem is the loose screws of the collimator holder (we didn't have the set screw so used a flat head screw). It's now fixed. 

We noticed another resonance peak near fundamental resonances. It's likely to be a p-pol. We will examine the polarizer and waveplates in the green input path. 

The home position of the PI stage controller was also changed after power outage. We'll re-reference the position after fixing the polarization. 

Images attached to this report
Displaying report 1-1 of 1.